Start-Up
Carl Zeiss Attaching colored disk carrier, polarizer or wide-angle unit Axiostar plus
2-20 B 40-815 e 12/01
1 Condenser carrier
2 Colored disk carrier
3 Stop bolt
4 Adjusting lever
5 Hexagonal key SW 2
6 Holding bolt
7 Slotted screwdriver
8 Wide angle unit or swiveling polarizer
Fig. 2-18 Attaching colored disk carrier, polarizer or
wide angle unit
2.9 Attaching colored disk carrier,
polarizer or wide-angle unit
The polarizer must be installed if you want to
use polarization contrast. The wide-angle unit
must be used with objectives having a
magnification < 4x so that the entire visible
field is illuminated.
• Hold colored disk carrier (2-18/2) parallel to
the underside of the condenser carrier
(2-18/1) and screw the holding bolts of the
colored disk carrier (2-18/6) into the front
thread on the bottom left of the condenser
carrier (2-18/1, position a) using the 90°-
offset hexagonal key SW 2 (2-18/5) until it
engages.
• Screw stop bolt (2-18/3) with adjusting lever
(2-18/4) into the rear thread on the
condenser carrier (2-18/1, position b) until it
engages.
• Screw wide-angle unit or swiveling polarizer
(2-18/8) with exceeding threaded bolt
(2-18/8, position a) into the front threaded
hole of the condenser carrier (2-18/1,
position a) by turning the entire wide-angle
unit (or the polarizer). Screw in the unit
until the threaded hole of the holding angle
(2-18/8 position b) lies below the rear
threaded hole (2-18/1, psition b) of the
condenser carrier and contacts the
underside of the condenser.
• Then use the slotted screwdriver to screw
the holding angle to the underside of the
condenser carrier (2-18/1, position b) with
the supplied slotted screw.