B 40 - 017 e / 12.96
5
4 Instructions for use
4.1 Adaptation of universal rotary stage
Axiolab Pol Axioplan Pol, Axioplan 2 Pol,
Axiophot Pol, Axiophot 2 Pol
Remove rotary stage in accordance with micro-
scope manual.
Center the rotary stage to the microscope in
accordance with the manual, take out the re-
ducing plate.
Remove objectives. Remove the Pol objective and Pol condenser.
Adjust the coarse drive to its upper limit stop.
Insert the 0.6 Pol UD condenser (Fig. 1/
2
) in the
condenser carrier of the universal rotary stage.
For this purpose, lower the condenser drive
(Fig. 3/
8
) as far as it will go and insert the UD
condenser (Fig. 1/
2
) in the dovetail guide of the
condenser carrier. Make sure the condenser is
correctly seated and clamp it (Fig. 3/
16
). Then
raise the condenser drive again (Fig. 3/
8
).
Move the stage carrier to its top stop, raise the
stage and lower the condenser carrier as far as
it will go.
Insert the 0.6 Pol UD condenser.(Fig. 1/
12
) (
Note!
Place the condenser in an almost vertical posi-
tion, slide it into the dovetail and clamp it, (see
microscope manual. If necessary, remove the
rotary stage from the stand, insert and clamp
the condenser and replace the rotary stage).
Then slighly move up the condenser drive.
Mount the universal rotary stage on the stand so
that the retaining pin (Fig. 3/
7
) is in contact with
the dovetail guide of the microscope and
clamp it (Fig. 3/
17
). The stage is now automati-
cally centered to the optical axis of the micro-
scope; lower the condenser carrier as far as it
will go.
Lower the stage carrier by about 15 mm and
clamp it. (See microscope manual.)
Place the UD 124 universal rotary stage on the
microscope stage (Fig. 4/
7
) and use the ball-
head screwdriver (Fig. 1/
8
) (or a 90° offset Allen
key) to tighten the two mounting screws (Fig. 4/
4
and
10
).
Set the tilt positions about the axes A
2
and A
4
to
0° click stops (Fig. 3/
1
and. Fig. 3/
19
), switch to
the click stops (Fig. 3/
13
and. Fig. 3/
14
), set the
graduated circle of the A
1
axis to the 0°-mark
(Fig. 3/
3
) and clamp all tilt movements (Fig. 3/
21
and
26
).
Set the tilt positions about the axes A
2
and A
4
to
0° click stops (Fig. 4/
1
and. Fig. 4/
15
), switch to
the click stops (Fig. 4/
13
and. Fig. 4/
14
set the
graduated circle of the A
1
axis to the 0°-mark
(Fig. 4/
3
) and clamp all tilt movements (Fig. 4/
17
and
22
)
Screw the UD Achromat Pol objectives (Fig. 1/
3
,
4
,
5
) into the nosepiece, always leaving one
nosepiece eye between each objective va-
cant. (It is recommended to use two objectives).
Use dust caps to close the vacant nosepiece
eyes.
Adjust the vertical condenser stop so that the
condenser will not hit the stage (see manual).
Check whether the spacer ring H“0“M 27 on
W 0.8 (4549109) has been screwed into the fixed
(not centerable) nosepiece eye. (Screw in if
necessary.) Screw the UD Achromat 20x/0.30 Pol
in this nose-piece eye, screw in the other objec-
tives so that one eye is left vacant between
each of them. Use dust caps to close the va-
cant nosepiece eyes.